EasyManuals Logo

Agilent Technologies 8753ET Service Guide

Agilent Technologies 8753ET
617 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #35 background imageLoading...
Page #35 background image
Chapter 2 2-9
System Verification and Performance Tests
Agilent 8753ES System Verification and Performance Tests
25.When the analyzer finishes measuring the standard, connect the 50 ohm termination to
reference test port 2.
26.Press .
27.When the analyzer finishes measuring the standard, press .
The analyzer briefly displays COMPUTING CAL COEFFICIENTS.
28.Connect the test port cables as shown Figure 2-3.
Figure 2-3 Transmission Calibration Setup
29.Press .
30.Press .
31.To save the calibration into the analyzer internal memory, press the following:
Rotate the knob to select Register 1 (REG1) on the display. Then press:
IMPORTANT Step 31 is crucial to the correct recall of the calibration during subsequent
measurements. The calibration MUST be stored in Register 1 (REG1) of
to be properly recalled.
32.When the analyzer finishes saving the instrument state, press
.
Device Verification
1. Press .
2. At the prompt, connect the 20 dB attenuator (supplied in the verification kit) as shown
in Figure 2-4.
REVERSE:LOAD
STANDARDS DONE
TRANSMISSION
DO BOTH FWD + REV
DONE 2-PORT CAL
Save/Recall
SELECT DISK
INTERNAL MEMORY
RETURN
SAVE STATE
INTERNAL MEMORY
SELECT DISK
INTERNAL DISK
System
SERVICE MENU
TESTS
28 x1
EXECUTE TEST

Table of Contents

Other manuals for Agilent Technologies 8753ET

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8753ET and is the answer not in the manual?

Agilent Technologies 8753ET Specifications

General IconGeneral
BrandAgilent Technologies
Model8753ET
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals