D Single-Mode/Multimode Module Performance Tests Test I. Dynamic Range
224 E6000C Mini-OTDR Users Guide, E0302
Settings 3 Set the OTDR:
[S
ETTINGS] menu:
–<R
ANGE> - select <RANGE INPUT...>:
Start: - enter value ST from Table 2, Table 3, or
Table 4. Confirm with <O
K>.
Span: - enter value SP from Table 2, Table 3, or Table
4. Confirm with <O
K>.
–<P
ULSE WIDTH>: enter value PW from Table 2, Table
3, or Table 4.
–<W
AVELENGTH>: If a dual wavelength module is
installed, select the required wavelength
–<M
EAS. MODE>: Averaging
–<O
PTIMIZE MODE>: Dynamic
–<A
VG. TIME>: 3 min
NOTE If the averaging parameter is listed for Number of
Averages, you should do the following:
Exit the [S
ETTINGS] menu
Press Ok.
Configuration •Enter the Instrument Config screen.
Select [C
ONFIG.]<INSTRUMENT CONFIG>
•Bring up the OTDR Settings page
Select [P
AGE INDEX]<OTDR SETTINGS>
Select Averaging time
Move to the Averaging Mode box and press select,
select Averaging time from the menu you see.
Save this configuration
Select Save.
Exit the Instrument Config screen
Select Ok.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E6000C and is the answer not in the manual?

Agilent Technologies E6000C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE6000C
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals