D Single-Mode/Multimode Module Performance Tests Test I. Dynamic Range
224 E6000C Mini-OTDR User’s Guide, E0302
Settings 3 Set the OTDR:
[S
ETTINGS] menu:
–<R
ANGE> - select <RANGE INPUT...>:
Start: - enter value ST from Table 2, Table 3, or
Table 4. Confirm with <O
K>.
Span: - enter value SP from Table 2, Table 3, or Table
4. Confirm with <O
K>.
–<P
ULSE WIDTH>: enter value PW from Table 2, Table
3, or Table 4.
–<W
AVELENGTH>: If a dual wavelength module is
installed, select the required wavelength
–<M
EAS. MODE>: Averaging
–<O
PTIMIZE MODE>: Dynamic
–<A
VG. TIME>: 3 min
NOTE If the averaging parameter is listed for Number of
Averages, you should do the following:
• Exit the [S
ETTINGS] menu
Press Ok.
Configuration •Enter the Instrument Config screen.
Select [C
ONFIG.]<INSTRUMENT CONFIG>
•Bring up the OTDR Settings page
Select [P
AGE INDEX]<OTDR SETTINGS>
• Select Averaging time
Move to the Averaging Mode box and press select,
select Averaging time from the menu you see.
• Save this configuration
Select Save.
• Exit the Instrument Config screen
Select Ok.