Page #133 background imageLoading...
Page #133 background image
CAPINTEC, INC CRC
®
-25R
Measure Tube # 5
1 + 4
Any Key to Continue
Figure 11-22 Lineator Measurement Prompt
Place the requested tube(s) in the Chamber and press any key (except HOME) to perform
the specified measurement.
Figure 11-23 Lineator Measurement Screen will appear.
Tc99m
6.27mCi
ENTER to Accept
Figure 11-23 Lineator Measurement Screen
Press ENTER to accept the measurement.
Figure 11-24 Lineator Measurement Verification
will appear.
Tube # 5
1 + 4
Present Factor:
n.nn
% Ratio: p.pp
OK? Y or N
Figure 11-24 Lineator Measurement Verification
To repeat the measurement or if the measurement is not acceptable, press NO.
Figure 11-22
Lineator Measurement Prompt
will re-appear.
If the measurement is acceptable, press YES. The next Lineator Measurement Prompt
screen will appear until all 8 tubes are measured.
After the last measurement, the results screens will appear.
August 07 ENHANCED TESTS 11 - 11

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Capintec CRC-25R and is the answer not in the manual?

Capintec CRC-25R Specifications

General IconGeneral
BrandCapintec
ModelCRC-25R
CategoryLaboratory Equipment
LanguageEnglish