CAPINTEC, INC CRC
®
-25R
ENTER ACTIVITY
value unit
_ mCi
-> to change unit
ENTER to Accept
Figure 12-18 Kit Activity Entry Screen
Input the value of the activity needed to make the kit and then press the RIGHT ARROW (Ô)
key. Use the arrow keys to scroll the activity units: mCi, µCi, Ci if using Curie mode, (or MBq,
GBq if using Becquerel mode). If the value needs to be changed, press CE.
When the activity value and units are correct, press ENTER. The kit input activity and volume
required for that activity will be displayed.
Withdraw the displayed volume and prepare the kit. Press any key (except HOME) to
continue and
Figure 12-19 Kit Measure Activity Screen will appear.
MEASURE
ACTIVITY
Any Key to Continue
Figure 12-19 Kit Measure Activity Screen
Place the kit in the Chamber and press any key (except HOME) to continue.
Figure 12-20
Kit Measurement Screen
will appear.
Tc99m
6.27mCi
ENTER to Accept
Figure 12-20 Kit Measurement Screen
When the activity has stabilized, press ENTER to accept the measurement.
Figure 12-21 Kit
Volume Entry Screen
will appear.
August 07 INVENTORY 12 - 9