v2.13 Cleverscope CS300 Reference Manual
©Cleverscope 2004-2018 www.cleverscope.com Page 115
contains up to 7 frequencies with corresponding Amp (Amplitude) values.
Enter values in ascending order. Overtype as necessary. You don't have to
fill in all entries. Just make sure that a pair of frequencies straddle the
frequency range you will be using.
interpolates from the first pair in the table to the last pair.
determines how the amplitude values are interpolated - log or linear.
Many amplifiers have a 20 dB/decade reducing gain characteristic and using
a log fill can compensate for this.
Here is an example plot using 4 values, sweeping from 0 to 200 kHz.
17.5.6 Measurements
This panel has been covered under the Frequency Analysis panel discussion. For Completeness here are the
panels.
Should not be used for PSU Input Impedance