EasyManua.ls Logo

Dexcom G5 - System Agreement to YSI Within CGM Glucose Ranges (Adult)

Dexcom G5
410 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Dexcom G5 Mobile System User Guide
302Technical Information
Table 1-A. System Agreement to YSI Within CGM Glucose Ranges (Adult)
CGM
Glucose
Range
1
(mg/dL)
Study
2
Number
of Paired
CGM-YSI
Percent
Within
15/15%
YSI
Percent
Within
20/20%
YSI
Percent
Within
30/30%
YSI
Percent
Greater
Than
40/40%
YSI
Overall
Original 9152 71% 82% 92% 3%
Software 505 2263 86% 93% 98% 1%
40-60
Original 512 67% 78% 88% 6%
Software 505 120 89% 94% 98% 0%
61-80
Original 781 73% 85% 94% 2%
Software 505 226 91% 96% 99% 0%
81-180
Original 3853 67% 78% 91% 3%
Software 505 738 84% 92% 98% 1%
181-300
Original 2784 72% 84% 93% 4%
Software 505 798 86% 93% 98% 1%
301-350
Original 775 82% 91% 97% 2%
Software 505 229 86% 94% 98% 1%
351-400
Original 447 74% 84% 91% 5%
Software 505 152 80% 92% 97% 0%
1
CGM readings are within 40-400 mg/dL, inclusive.
2
Both sets of study data are presented and are labeled as Original (SW10050) or Software
505
(SW10505).