Dexcom G5 Mobile System User Guide
302Technical Information
Table 1-A. System Agreement to YSI Within CGM Glucose Ranges (Adult)
CGM
Glucose
Range
1
(mg/dL)
Study
2
Number
of Paired
CGM-YSI
Percent
Within
15/15%
YSI
Percent
Within
20/20%
YSI
Percent
Within
30/30%
YSI
Percent
Greater
Than
40/40%
YSI
Overall
Original 9152 71% 82% 92% 3%
Software 505 2263 86% 93% 98% 1%
40-60
Original 512 67% 78% 88% 6%
Software 505 120 89% 94% 98% 0%
61-80
Original 781 73% 85% 94% 2%
Software 505 226 91% 96% 99% 0%
81-180
Original 3853 67% 78% 91% 3%
Software 505 738 84% 92% 98% 1%
181-300
Original 2784 72% 84% 93% 4%
Software 505 798 86% 93% 98% 1%
301-350
Original 775 82% 91% 97% 2%
Software 505 229 86% 94% 98% 1%
351-400
Original 447 74% 84% 91% 5%
Software 505 152 80% 92% 97% 0%
1
CGM readings are within 40-400 mg/dL, inclusive.
2
Both sets of study data are presented and are labeled as Original (SW10050) or Software
505
(SW10505).