334Technical Information
Table 11-B. System Agreement to SMBG Within CGM Glucose Ranges
(Pediatric)
CGM
Glucose
Range
1
(mg/dL)
Study
2
Number
of Paired
CGM-SMBG
Percent
Within
15/15%
SMBG
Percent
Within
20/20%
SMBG
Percent
Within
30/30%
SMBG
Percent
Greater
Than
40/40%
SMBG
Overall
Original 16,318 64% 76% 89% 5%
Software 505 4264 73% 84% 94% 2%
40-60
Original 487 44% 55% 68% 19%
Software 505 240 54% 71% 86% 7%
61-80
Original 1340 59% 70% 85% 7%
Software 505 399 64% 76% 92% 2%
81-180
Original 7084 62% 74% 90% 5%
Software 505 1650 72% 84% 95% 2%
181-300
Original 5627 69% 80% 90% 5%
Software 505 1526 79% 89% 97% 2%
301-350
Original 1176 65% 77% 90% 4%
Software 505 319 72% 83% 94% 2%
351-400
Original 604 58% 72% 86% 6%
Software 505 130 69% 79% 86% 8%
1
CGM readings are within 40 to 400 mg/dL, inclusive.
2
Both sets of study data are presented and are labeled as Original (SW10050) or Software
505