EasyManua.ls Logo

Dexcom G5 - System Agreement to SMBG Within CGM Glucose Ranges (Pediatric)

Dexcom G5
410 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Dexcom G5 Mobile System User Guide
334Technical Information
Table 11-B. System Agreement to SMBG Within CGM Glucose Ranges
(Pediatric)
CGM
Glucose
Range
1
(mg/dL)
Study
2
Number
of Paired
CGM-SMBG
Percent
Within
15/15%
SMBG
Percent
Within
20/20%
SMBG
Percent
Within
30/30%
SMBG
Percent
Greater
Than
40/40%
SMBG
Overall
Original 16,318 64% 76% 89% 5%
Software 505 4264 73% 84% 94% 2%
40-60
Original 487 44% 55% 68% 19%
Software 505 240 54% 71% 86% 7%
61-80
Original 1340 59% 70% 85% 7%
Software 505 399 64% 76% 92% 2%
81-180
Original 7084 62% 74% 90% 5%
Software 505 1650 72% 84% 95% 2%
181-300
Original 5627 69% 80% 90% 5%
Software 505 1526 79% 89% 97% 2%
301-350
Original 1176 65% 77% 90% 4%
Software 505 319 72% 83% 94% 2%
351-400
Original 604 58% 72% 86% 6%
Software 505 130 69% 79% 86% 8%
1
CGM readings are within 40 to 400 mg/dL, inclusive.
2
Both sets of study data are presented and are labeled as Original (SW10050) or Software
505
(SW10505).

Table of Contents

Related product manuals