EasyManua.ls Logo

Dexcom G5 - System Agreement to SMBG Within CGM Glucose Ranges (Adult)

Dexcom G5
410 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Dexcom G5 Mobile System User Guide
333Technical Information
Table 11-A. System Agreement to SMBG Within CGM Glucose Ranges
(Adult)
CGM
Glucose
Range
1
(mg/dL)
Study
2
Number
of Paired
CGM-SMBG
Percent
Within
15/15%
SMBG
Percent
Within
20/20%
SMBG
Percent
Within
30/30%
SMBG
Percent
Greater
Than
40/40%
SMBG
Overall
Original 7508 69% 81% 94% 2%
Software 505 2992 77% 87% 96% 1%
40-60
Original 731 75% 84% 92% 4%
Software 505 221 73% 80% 87% 7%
61-80
Original 968 78% 86% 95% 1%
Software 505 336 77% 85% 95% 1%
81-180
Original 3141 65% 78% 93% 2%
Software 505 1362 74% 85% 96% 1%
181-300
Original 1960 68% 81% 94% 3%
Software 505 826 80% 90% 97% 1%
301-350
Original 450 77% 88% 98% 1%
Software 505 161 83% 93% 99% 0%
351-400
Original 258 75% 85% 95% 2%
Software 505 86 90% 93% 98% 1%
1
CGM readings are within 40 to 400 mg/dL, inclusive.
2
Both sets of study data are presented and are labeled as Original (SW10050) or Software
505
(SW10505).

Table of Contents

Related product manuals