Dexcom G5 Mobile System User Guide
321Technical Information
Table 6-C. Hyperglycemia Alert and Detection Rate Evaluation in
Reference to SMBG 30 Minutes Before and After (Pediatric, Ages 2-5
Years)
Hyperglycemia
Alert Level
(mg/dL)
Study
1
True
Alert
Rate
False
Alert
Rate
Hyperglycemia
Detection Rate
Hyperglycemia
Missed
Detection Rate
120
Original 92% 8% 98% 2%
Software 505 97% 3% 99% 1%
140
Original 90% 10% 98% 2%
Software 505 98% 2% 100% 0%
180
Original 87% 13% 96% 4%
Software 505 99% 1% 93% 7%
200
Original 85% 15% 96% 4%
Software 505 98% 2% 93% 7%
220
Original 81% 19% 95% 5%
Software 505 100% 0% 97% 3%
240
Original 80% 20% 95% 5%
Software 505 99% 1% 98% 2%
300
Original 71% 29% 90% 10%
Software 505 95% 5% 96% 4%
1
Both sets of study data are presented and are labeled as Original (SW10050) or Software
505
(SW10505).
The High Glucose Alert
Estimates of how well the adjustable High Glucose Alert performs are presented in Table 6-A, 6-B and
6-C. Table 6-A represents the hyperglycemia alert evaluation within 15 minutes of the YSI value in the
adult studies. Table 6-B represents the Alert evaluation within 15 minutes of the YSI value for a subset
of the pediatric populationsubjects age 6 to 17 years who had YSI measurements every 15 minutes.
Table 6-C represents the Alert evaluation within 30 minutes of an SMBG reading for 2- to 5-year-old
subjects in the pediatric studies.