Dexcom G5 Mobile System User Guide
316Technical Information
Table 5-B. Hypoglycemia Alert and Detection Rate Evaluation in
Reference to YSI 15 Minutes Before and After (Pediatric, Ages 6-17 Years)
Hypoglycemia
Alert Level
(mg/dL)
Study
1
True
Alert
Rate
False
Alert
Rate
Hypoglycemia
Detection Rate
Hypoglycemia
Missed
Detection Rate
55
Original 0% 100% 0% 100%
Software 505 22% 78% 75% 25%
60
Original 11% 89% 25% 75%
Software 505 42% 58% 78% 23%
70
Original 47% 53% 50% 50%
Software 505 68% 32% 75% 25%
80
Original 55% 45% 55% 45%
Software 505 86% 14% 91% 9%
90
Original 69% 31% 62% 38%
Software 505 90% 10% 93% 7%
100
Original 75% 25% 62% 38%
Software 505 91% 9% 93% 7%
1
Both sets of study data are presented and are labeled as Original (SW10050) or Software
505
(SW10505).