Dexcom G5 Mobile System User Guide
310Technical Information
Table 3-C. Concurrence of CGM Readings and YSI Values (Original
Pediatric Study)
CGM
(mg/dL)
YSI (mg/dL)
Row Percentage of Matched Pairs in Each CGM Glucose Range
< 40
40-
60
61-
80
81-
120
121-
160
161-
200
201-
250
251-
300
301-
350
351-
400
>
400
Number
of Paired
CGM-YSI
< 40 0% 0% 0% 54% 31% 15% 0% 0% 0% 0% 0% 13
40-60 0% 21% 58% 16% 5% 0% 0% 0% 0% 0% 0% 19
61-80 0% 21% 45% 30% 4% 0% 0% 0% 0% 0% 0% 76
81-120 0% 1% 20% 66% 12% 1% 0% 0% 0% 0% 0% 338
121-160 0% 0% 1% 36% 54% 7% 1% 0% 0% 0% 0% 511
161-200 0% 0% 0% 4% 40% 48% 6% 1% 0% 0% 0% 596
201-250 0% 0% 0% 1% 9% 44% 41% 5% 0% 0% 0% 658
251-300 0% 0% 0% 0% 2% 7% 50% 36% 3% 0% 2% 432
301-350 0% 0% 0% 0% 0% 2% 18% 59% 21% 0% 0% 206
351-400 0% 0% 0% 0% 0% 0% 3% 28% 50% 16% 2% 86
> 400 0% 0% 0% 0% 0% 0% 1% 14% 41% 36% 7% 70