EasyManua.ls Logo

Dexcom G5 - Concurrence of CGM Readings and YSI Values (Original Adult Study)

Dexcom G5
410 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Dexcom G5 Mobile System User Guide
308Technical Information
Table 3-A. Concurrence of CGM Readings and YSI Values (Original Adult
Study)
CGM
(mg/dL)
YSI (mg/dL)
Row Percentage of Matched Pairs in Each CGM Glucose Range
< 40
40-
60
61-
80
81-
120
121-
160
161-
200
201-
250
251-
300
301-
350
351-
400
>
400
Number
of Paired
CGM-YSI
< 40 6% 48% 37% 7% 1% 0% 0% 0% 0% 0% 0% 155
40-60 4% 49% 36% 11% 1% 0% 0% 0% 0% 0% 0% 512
61-80 0% 22% 51% 24% 1% 0% 0% 0% 0% 0% 0% 781
81-120 0% 2% 17% 66% 13% 1% 0% 0% 0% 0% 0% 1706
121-160 0% 0% 1% 25% 60% 13% 2% 0% 0% 0% 0% 1492
161-200 0% 0% 0% 2% 28% 53% 16% 2% 0% 0% 0% 1240
201-250 0% 0% 0% 0% 3% 21% 51% 21% 3% 1% 0% 1181
251-300 0% 0% 0% 0% 0% 4% 19% 49% 24% 3% 0% 1018
301-350 0% 0% 0% 0% 0% 0% 3% 28% 51% 16% 1% 775
351-400 0% 0% 0% 0% 0% 0% 3% 10% 43% 38% 7% 447
> 400 0% 0% 0% 0% 0% 0% 1% 6% 21% 57% 15% 248