Dexcom G5 Mobile System User Guide
320Technical Information
Table 6-B. Hyperglycemia Alert and Detection Rate Evaluation in
Reference to YSI 15 Minutes Before and After (Pediatric, Ages 6-17 Years)
Hyperglycemia
Alert Level
(mg/dL)
Study
1
True
Alert
Rate
False
Alert
Rate
Hyperglycemia
Detection Rate
Hyperglycemia
Missed
Detection Rate
120
Original 91% 9% 98% 2%
Software 505 98% 2% 99% 1%
140
Original 87% 13% 99% 1%
Software 505 97% 3% 98% 2%
180
Original 75% 25% 99% 1%
Software 505 94% 6% 98% 2%
200
Original 71% 29% 98% 2%
Software 505 94% 6% 97% 3%
220
Original 67% 33% 97% 3%
Software 505 93% 7% 96% 4%
240
Original 62% 38% 96% 4%
Software 505 88% 12% 94% 6%
300
Original 43% 57% 93% 7%
Software 505 69% 31% 84% 16%
1
Both sets of study data are presented and are labeled as Original (SW10050) or Software
505
(SW10505).