Dexcom G5 Mobile System User Guide
331Technical Information
Table 10-A. System Readings Within Wear Days (Adult)
Statistic Study
1
Day 1 Day 2 Day 3 Day 4 Day 5 Day 6 Day 7
All
Days
2
Mean
Original 98% 98% 98% 98% 97% 99% 95% 97%
Software 505 98% 99% 98% 98% 96% 99% 97% 98%
Median
Original 100% 100% 100% 100% 100% 100% 100% 100%
Software 505 99% 100% 100% 100% 100% 100% 100% 100%
Standard
Deviation
Original 5% 3% 9% 8% 10% 3% 11% 8%
Software 505 3% 2% 8% 11% 15% 2% 13% 9%
1
Both sets of study data are presented and are labeled as Original (SW10050) or Software
505
(SW10505)
.
2
A total of 108 sensors were included with the Original Study and 51 sensors were included with the
Software 505 Study.