Dexcom G5 Mobile System User Guide
315Technical Information
Table 5-A. Hypoglycemia Alert and Detection Rate Evaluation in
Reference to YSI 15 Minutes Before and After (Adult)
Hypoglycemia
Alert Level
(mg/dL)
Study
1
True
Alert
Rate
False
Alert
Rate
Hypoglycemia
Detection Rate
Hypoglycemia
Missed
Detection Rate
55
Original 50% 50% 71% 29%
Software 505 71% 29% 68% 32%
60
Original 64% 36% 75% 25%
Software 505 85% 15% 83% 17%
70
Original 79% 21% 83% 17%
Software 505 92% 8% 91% 9%
80
Original 87% 13% 86% 14%
Software 505 95% 5% 90% 10%
90
Original 90% 10% 89% 11%
Software 505 96% 4% 94% 6%
1