EasyManua.ls Logo

Dexcom G5 - Sensor Stability Relative to YSI (Accuracy over Time) - (Adult)

Dexcom G5
410 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Dexcom G5 Mobile System User Guide
325Technical Information
Table 8-A. Sensor Stability Relative to YSI (Accuracy Over Time
1
) - (Adult)
Day
of
Wear
Study
2
Number
of
Paired
CGM-
YSI
Mean
Absolute
Percent
Differences
Median
Absolute
Percent
Differences
Percent
Within
15/15%
YSI
Percent
Within
20/20%
YSI
Percent
Within
30/30%
YSI
Percent
Greater
Than
40/40%
YSI
Day
1
Original 3023 16.7% 13.2% 59% 71% 86% 6%
Software 505 680 10.7% 7.9% 77% 84% 96% 2%
Day
4
Original 3108 11.4% 8.2% 77% 87% 95% 2%
Software 505 777 8.0% 6.4% 89% 96% 99% 0%
Day
7
Original 3021 11.9% 8.9% 76% 87% 95% 2%
Software 505 806 8.5% 7.2% 90% 97% 99% 0%
1
CGM readings are within 40 to 400 mg/dL, inclusive.
2
Both sets of study data are presented and are labeled as Original (SW10050) or Software
505
(SW10505).

Table of Contents