Dexcom G5 Mobile System User Guide
305Technical Information
Table 2-A. Number and Percentage of YSI Values When CGM Readings
Are “LOW” or “HIGH” (Adult)
YSI mg/dL
Total
CGM
Readings
Study
1
CGM-YSI
Pairs
< 55 < 60 < 70 < 80 ≥ 80
“LOW
Original
n 66 84 123 142 13 155
Cumulative
Percent
42% 54% 79% 92% 8%
Software
505
n 11 16 17 18 0 18
Cumulative
Percent
61% 89% 94% 100% 0%
YSI mg/dL
Total
CGM
Readings
Study
1
CGM-YSI
Pairs
> 340 > 320 > 280 > 240 240
“HIGH”
Original
n 189 220 238 246 2 248
Cumulative
Percent
76% 89% 96% 99% 1%
Software
505
n 40 43 45 45 0 45
Cumulative
Percent
89% 96% 100% 100% 0%
1
Both sets of study data are presented and are labeled as Original (SW10050) or Software
505
(SW10505).