307Technical Information
For example, in the Software 505 Adult Study (Table 2-A), when the system displayed “LOW” (18
occasions), 100% (18 out of 18) of the YSI values were less than 80 mg/dL, and 94% (17 out of 18) of
the YSI values were less than 70 mg/dL. When the system displayed “HIGH” (45 occasions), 100% (45
out of 45) of the YSI values were greater than 240 mg/dL, and 100% (45 out of 45) of the YSI values
were greater than 280 mg/dL.