Dexcom G5 Mobile System User Guide
97Calibration
PRECAUTION
Do:
Look at trend arrows before calibrating. Trend arrows help you determine if you can
calibrate now or should wait.
Don’t:
Never calibrate if your BG is changing at a significant rate, typically more than 2
mg/dL per minute.
Never calibrate when you see:
A single arrow, pointing up
Rising 2-3 mg/dL each minute
Two arrows pointing up
Rising more than 3 mg/dL each minute
Single arrow pointing down
Falling 2-3 mg/dL each minute
Two arrows pointing down
Falling more than 3 mg/dL each minute
Why: Calibrating during significant rise/fall of BG may affect accuracy of sensor glucose
readings.
Consequences: Missing a severe low or high glucose event or making a treatment
decision that results in injury.
PRECAUTION
Do:
Enter the exact BG value displayed on your BG meter within five minutes of a carefully
performed fingerstick measurement.
Why: Entering the wrong BG values, or waiting more than five minutes before entry, might
affect sensor accuracy.
Consequences: Missing a severe low or high glucose event or making a treatment
decision that results in injury.