OVERVIEW
VF6 Infinity Vista D-5
AC Power Adapter (59 53 539 E530U)
Power Requirements 100-120 VAC, 0.8A
200-240 VAC, 0.4A
Output Max. 34 W
Frequency 50-60 Hz
Chassis Leakage Current d300 PA @ 110 V and 60 Hz (UL 544)
d500 PA @ 220 V and 50 Hz (IEC 601-1)
Mode of Operation Continuous
Protection Class Class I
R50 N Network Recorder
Dimensions 180 x 120 x 222 mm (7.1 x 4.72 x 8.74 in.)
Weight 1.64 kg (3.6 lb)
Connections AC Power Connector, Infinity Network, R50 Recorder,
Potential Equalization Connector.
Cooling Convection
Input voltage range 100-240 VRMS
Mains frequency 50/60 Hz
Power Consumption 1.0 A max
Protection Class Class I
Chassis Leakage Current <300 mA @ 120VAC, <500 mA @ 220VAC
Mode of Operation Continuous
Water Resistance Ordinary
Environmental Requirements Temperature:
Operating: 15°C to 40°C (55°F to 104°F)
Storage: -20°C to 40°C (-4°F to 104°F)
Relative Humidity:
Operating: 30% to 95%, non-condensing
Storage: 10% to 95%, non-condensing
with packaging
Atmospheric Pressure:
Operating: 550 to 775 mmHg (73 to 103 kPa)
Storage: 375 to 795 mmHg (50 to 106 kPa)