X-STREAM
Emerson Process Management GmbH & Co. OHG6-14
Instruction Manual
HASAxE-IM-HS
05/2006
6-2 Checking the Instrument's Setup
Note!
"Function" parameter factory setting is Off FS
(fail-safe) unless otherwise specified at time
of order.
Combining the different modes with setting
the threshold levels in a proper way gives
the operator the choice of programming diffe-
rent behaviours:
• Window mode: If the concentration
exceeds the limits defining a concentra-
tion window, an alarm is activated.
• High and high-high alarms: This mode
allows to set a prealarm level and a main
alarm level for increasing concentrations
• Low and low-low alarms: This mode allows
to set a prealarm and a main alarm for
decreasing concentrations.
Refer to the following paragraphs and figures
for detailed information on alarm settings!
Each threshold may be configured to be OFF,
a LOW or a HIGH alarm, utilizing the "Function"
parameter:
Off: Alarm function deactivated, output level
is digital "LOW" all the time.
Low: The moment, the currently measured
concentration goes below the specified
level, an alarm signal is activated.
Alarm signal level is digital "HIGH".
High: The moment, the currently measured
concentration goes above the specified
level, an alarm signal is activated.
Alarm signal level is digital "HIGH".
In addition the "Function" parameter supports
the fail-safe operation mode:
Fail-Safe means an activated alarm gives a
digital "LOW" level signal instead of a "HIGH"
level signal:
Off FS: Alarm function deactivated, output level
is digital "HIGH" all the time.
Low FS: The moment, the currently measured
concentration goes below the specified
level, an alarm signal is activated.
Alarm signal level is digital "LOW".
High FS: The moment, the currently measured
concentration goes above the specified
level, an alarm signal is activated.
Alarm signal level is digital "LOW".

Table of Contents

Other manuals for Emerson Rosemount Xstream Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Rosemount Xstream Series and is the answer not in the manual?

Emerson Rosemount Xstream Series Specifications

General IconGeneral
BrandEmerson
ModelRosemount Xstream Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals