X-STREAM
Emerson Process Management GmbH & Co. OHG8-16
Instruction Manual
HASAxE-IM-HS
05/2006
8-4 Troubleshooting on Components
8-4-3-1-6 Analog Preamplifier
a) Paramagnetic Oxygen Measurement
Measuring Point: X 25 (channel 1)
Measuring instrument: DVM
Signal when connected to
zero gas: 0 V dc (± 50 mV)
ambient air (approx. 21 Vol. - % O
2
):
100 % O2 sensor: approx. 840 mV
25 % O2 sensor: approx. 3.36 V
(Type of sensor: see separate nameplate label)
Failure: No signal or wrong measuring values
Potential cause:
a) Oxygen sensor not connected
b) Oxygen sensor defective
c) BKS board defective
How to solve:
a) Connect Oxygen sensor
b) Replace Oxygen sensor
c) Replace BKS board
b) IR Measurement
Measuring Point:
X 25 (channel 1; not for instruments with
Oxygen measurement)
X 27 (channel 2)
Measuring instrument: DVM
Signal when connected to
zero gas: 0 V dc (± 100 mV)
Zero point voltage and span voltage must differ
at least by 600 mV (at ranges < 1000 ppm
difference should be 500 mV min.)
Failure: No signal or wrong measuring values
Potential cause:
a) Detector not connected
b) Detector defective
c) BKS board defective
How to solve:
a) Connect detector
b) Replace detector
c) Replace BKS board

Table of Contents

Other manuals for Emerson Rosemount Xstream Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Rosemount Xstream Series and is the answer not in the manual?

Emerson Rosemount Xstream Series Specifications

General IconGeneral
BrandEmerson
ModelRosemount Xstream Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals