X-STREAM
3-17
Instruction Manual
HASAxE-IM-HS
05/2006
Emerson Process Management GmbH & Co. OHG
3 Principles
3-4 Measurement Specification
Performance Specifications
Table 3-7: Measurement Performance Specifications
1
Related to full scale
2
Related to measuring value;
1 psi = 68.95 hPa
3
From gas analyzer inlet at 1.0 l/min gas
flow (electr. = 2 s)
4
Constant pressure and temperature
5
Dependent on integrated photometer bench
6
Paramagnetical oxygen measurement (pO
2
)
NDIR/UV/VIS Oxygen Sensor (pO
2
and eO
2
) Thermal Conductivity
Detection limit
<
1%
1 4
<
1%
1 4
<
2%
1 4
Linearity
<
1%
1 4
<
1%
1 4
<
1%
1 4
Zero-point drift
<
2% per week
1 4
<
2% per week
1 4
<
2% per week
1 4
Span (sensitivity) drift
<
1% per week
1 4
<
1% per week
1
<
1% per week
1 4
Repeatability
<
1%
1 4
<
1%
1 4
<
1%
1 4
Response time (t
90
) 4 s
<
t
90
<
7 s
3 5
< 5 s
3 6
/ approx. 12 s
3 9
5 s
<
t
90
<
20 s
3 7
Permissible gas flow 0.2 - 1.5 l/min. 0.2 - 1.0 l/min
6
/ 0.2 - 1.5 l/min.
9
0.2 - 1.5 l/min. (
+
0.1 l/min)
Influence of gas flow
<
0.5%
1
4
<
2%
1
4
<
1%
1 4 13
Maximum gas pressure
<
1,500 hPa abs. (
<
7 psig) Atm. pressure
6
/
<
1,500 hPa abs.
9
(
<
7 psig)
<
1,500 hPa abs. (< 7 psig)
Influence of pressure
- At constant temperature
<
0.10% per hPa
2
< 0.10 % per hPa
2
<
0.10% per hPa
2
- With pressure compensation
8
<
0.01% per hPa
2
<
0.01 % per hPa
2
<
0.01% per hPa
2
Permissible ambient temperature -20 to +50 °C (-4 to +122 °F) -20 to +50 °C (-4 to +122 °F)
10
-20 to +50 °C(-4 to +122 °F)
Influence of temperature
(at constant pressure)
- On zero point
<
1% per 10 K
1
<
1% per 10 K
1
< 1% per 10 K
1 15
- On span (sensitivity)
<
5% (0 to +50 °C)
1 11 15
<
1% per 10 K
1 15
<
1% per 10 K
1 15
Thermostat control
12 14
Optionally 60 °C (140 °F) 55/60 °C (131/140 °F)
6
/ None
9
75 °C (167 °F)
12
Warm-up time
12 14
15 to 50 minutes
5
Approx. 50 minutes
6
Approx. 15 minutes
12
Sensor / cell only
13
Flow variation within
+
0.1 l/min
14
Option "thermostated box" with temperature
60 °C (140 °F)
15
Temperature variation; 10 K in 1 h
7
Depending on measuring range
8
Pressure sensor is required
9
Electrochemical oxygen measurement (eO
2
), not for use
with sample gas containing FCHC´s
10
Electrochemical oxygen measurement (eO
2
):
+5 to +40 °C (41 to 104 °F)
11
Starting from 20°C/68 °F (to 0 °C / 32 °F and
to +50 °C / 122 °F)

Table of Contents

Other manuals for Emerson Rosemount Xstream Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Rosemount Xstream Series and is the answer not in the manual?

Emerson Rosemount Xstream Series Specifications

General IconGeneral
BrandEmerson
ModelRosemount Xstream Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals