X-STREAM
4-14
Instruction Manual
HASAxE-IM-HS
07/2006
Emerson Process Management GmbH & Co. OHG
All signal cables are to be connected via screw
terminals, located at the analyzer's rear panel.
The strain reliefs provide a metal strip to connect
Preparation of signal cables
Supported wire cross sections: 24 to 14 AWG (0.2 to 2.5 mm
2
),
no need to use wire end sleeves
Cable skinning length: 0.354 inch (9 mm)
Hole diameter: 0.05 inch (1.2 mm)
Screw thread: M 2.5
Tightening torque, min: 3.5 in.lb (0.4 Nm)
Signal inputs / outputs
Analog signal outputs
To connect the terminals remove the cover at
the analyzer's rear panel (4 screws). Feed the
analog signals cable through the uppermost
edge protection, and through the upper strain-
relief.
The upper 4 terminals (# 1 - 4) of the terminals
row next to the power terminals are reserved
for analog signal outputs.
4-4-1 Installation - X-STREAM GP, X-STREAM GPS
to the cable shield after the cable's outer
insulation is stripped.
Fig. 4-11: Analog Signal Output Terminals
(+) 4 (0) - 20 mA, channel 1
(-) 4 (0) - 20 mA, channel 1
(+) 4 (0) - 20 mA, channel 2
(-) 4 (0) - 20 mA, channel 2
Note!
Take care of the special installation
instructions in section 4-5!
Interface specification:
5-4-3-4-1 Analog output setup, page 5-31
Burden: R
B
≤ 500 Ω

Table of Contents

Other manuals for Emerson Rosemount Xstream Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Rosemount Xstream Series and is the answer not in the manual?

Emerson Rosemount Xstream Series Specifications

General IconGeneral
BrandEmerson
ModelRosemount Xstream Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals