X-STREAM
5-28
Instruction Manual
HASAxE-IM-HS
05/2006
Emerson Process Management GmbH & Co. OHG
5-4 Menu System - Calibration Setup Menu
5-4-3-2-2 Interval Time Setup Menu
Enter the time between two zero calibrations
when in autocal mode.
If system is setup accordingly access level 3
code must be entered to access this menu.
Enter the time between two span calibrations
when in autocal mode.
Input range for both values: 0 .. 999 h
Note!
This menu is available only if the parameter
"Valves" in the INSTALLED OPTIONS setup
menu is other than none
Setup..
Calibration..
Interval time..
Interval time
ZeroCal: 2 h
SpanCal: 10 h
Note!
For detailed descriptions of how to perform
calibrations chapter 7 Maintenance.