X-STREAM
4-26
Instruction Manual
HASAxE-IM-HS
07/2006
Emerson Process Management GmbH & Co. OHG
4-4-2 Installation - X-STREAM F
Digital Inputs & Outputs
For accessing the terminals open the
instrument's front door. The leftmost terminal
strip is reserved for the digital inputs and
outputs.
Fig. 4-23: Digital Input & Output Terminals
Design: Open collector (outputs)
Electrical specification:
outputs: max. 30 V , 30 mA
inputs: max. 30 V , internally limited to 2.3 mA
H level: min. 4 V; L level: max. 3 V
Note!
Take care of the special installation instruc-
tions in section 4-5 and of the cable gland
assembly instruction on page 4-22!
Ch1: threshold 1
Ch1: threshold 2
Ch2: threshold 1
Ch2: threshold 2
Sample valve
Valve V4
Valve V1
Valve V2
DIG Out GND
Zero cal. Ch1 & Ch2
Span cal. Ch1
Span cal. Ch2
Open V4
Open V1
Open V2
Open sample valve or close all
DIG In GND
Terminal 1
Terminal 2
Terminal 3
Terminal 4
Terminal 5
Terminal 6
Terminal 7
Terminal 8
Terminal 9
Terminal 10
Terminal 11
Terminal 12
Terminal 13
Terminal 14
Terminal 15
Terminal 16
Terminal 17
Digital Inputs DigitalOutputs

Table of Contents

Other manuals for Emerson Rosemount Xstream Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Rosemount Xstream Series and is the answer not in the manual?

Emerson Rosemount Xstream Series Specifications

General IconGeneral
BrandEmerson
ModelRosemount Xstream Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals