X-STREAM
2-6
Instruction Manual
HASAxE-IM-HS
05/2006
Emerson Process Management GmbH & Co. OHG
Housing
Permissible ambient temperature range 32 F to 122 F ( 0 °C to +50 °C)
Weight: approx. 26.5 to 35.3 lbs (12 - 16 kg)
(depending on analyzer configuration)
Protection class to EN 60529: IP 20 for indoor installation
Analyzer must not be exposed to dripping
or sprayed water.
Gas fittings: quantity: max. 4
specification: 6/4 mm PVDF
optional 6/4 mm or 1/4", stainless
steel, other on request
2-2-1 X-STREAM GP, X-STREAM GPS
2-2-1-1 Terminals Version
Power connection
Connection via covered screw terminals located
at the analyzer's rear panel (fig. 2-2).
Cross section: max. 12 AWG (2.5 mm
2
), using conductor
sleeves is not required
Cable entry through cover via 1 strain relief
Power fuses
Two fuses are located nearby the power terminals.
Fuse ratings: AC 230 V, 3.15 A, 5x20 mm

Table of Contents

Other manuals for Emerson Rosemount Xstream Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Rosemount Xstream Series and is the answer not in the manual?

Emerson Rosemount Xstream Series Specifications

General IconGeneral
BrandEmerson
ModelRosemount Xstream Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals