X-STREAM
4-4
Instruction Manual
HASAxE-IM-HS
07/2006
Emerson Process Management GmbH & Co. OHG
Internal Solenoid Valve Block
Supply overpressure for all gases is limited to
0.7 to 7.2 psi (50 to 500 hPa) when the analyzer
is equipped with an internal solenoid valve
block.
Overpressure valve
Flowmeter
Output
Output
Analyzer
Sample pump
Filter
Fig. 4-1: Bypass Mode Installation
The analyzer should be mounted near the
sample source to minimize sample transport
time. A sample pump may be used to decrease
response time, whereat the analyzer is either
operated in bypass mode or protected by an
overpressure valve against too high flow and
pressure (fig. 4-1).
4-2 Installation - Gas Conditioning