X-STREAM
Emerson Process Management GmbH & Co. OHG 9-3
Instruction Manual
HASAxE-IM-HS
05/2006
9 Modbus
This list is ordered alphabetically by the parameter name („Mnemonic“) column.
Any "#" is a wildcard, to be replaced by "1" or "2" for the desired measuring channel. In this case
the upper register line is assigned to channel 1, the lower line to channel 2.
9-3 List of Parameters and Registers
9-3 Modbus - List of Parameters and Registers
Attribute
1
Daniel /
Enron
Modicon Type
3066
3067
3066
3067
ActivateSimulValue#DB
activates simulation of concentration
measurement values
bit0: RawValueConcentration#DB
bit1: ZeroOffset#DB
bit2: TemperatureOffset#DB
bit3: CrossInterferenceOffset#DB
bit4: TemperatureFactor#DB
bit5: PressureFactor#DB
bit6: MEASx
bit7: Concentration#DB
Par, Fac
Byte-Bitfield
3131..3138 3131..3138
ActivSimulMuxValuesDB
activates simulation of the multiplexed values
of the 3rd AD-converter
bit0: ADC-value
bit1: averaged ADC-value
bit2: reserved
bit3: reserved
bit4: reserved
bit5: reserved
bit6: reserved
bit7: finale value
Par, Fac
Byte-Bitfield
-
Arr8
9081..9082
9083..9084
6161..6164
6165..6168
ADC_Isrc#CoeffsDB
polynom coefficients that convert ADC-value
into current of light source1 in mA
A0, A1
Cfg, Fac
Float-Arr2
9041..9043 6081..6086
ADC_Temp1CoeffsDB
polynom coefficients that convert ADC-value
into Temperature1 in °C
A0, A1, A2
Cfg, Fac
Float-Arr3
9044..9046 6087..6092
ADC_Temp2CoeffsDB
polynom coefficients that convert ADC-value
into Temperature2 in °C
A0, A1, A2
Cfg, Fac
Float-Arr3
1001
1002
1001
1002
AdConversionError#DB
raw measurement of concentration is not
running
Dyn
Boolean
1003 1003
AdConversionTemperatureDB
raw measurement of 3rd AD-converter is not
running
Dyn
Boolean
3103
3104
3103
3104
ADconvOffset#DB
defines the AD-converter# correction offset
Cfg, Fac
Byte
3102 3102
ADconvOffsetMuxDB
defines the AD-converter3(Mux) correction
offset
Cfg, Fac
Byte
9025 6049..6050
AirPressureDB
current air pressure (in hPa):
if internal pressure measurement is enabled
(PressureSensorInstalledDB) this is a dynamic
variable;
if no pressure sensor is installed we have to
input the current value
500 .. 2000
Dyn, Cfg
Float
Range / Enum values / Coefficients
Modbus-Register
Mnemonic
Description

Table of Contents

Other manuals for Emerson Rosemount Xstream Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Rosemount Xstream Series and is the answer not in the manual?

Emerson Rosemount Xstream Series Specifications

General IconGeneral
BrandEmerson
ModelRosemount Xstream Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals