CR Console Operation Manual
2008. 09 897N0519F 4 - 3
(3) Using Utility
Operates various utility functions (Magnifying display image/Changing parameters/Reprinting
display image/Setting rejected image/Locking study/Saving image/Outputting DICOM files/Image
special print/Resetting).
Perform these operations accordingly. The “Magnifying display image*
1
/Changing
parameters/Reprinting display image/Saving image/Resetting” are common functions and can be
used at the Image Edit Screen by an image level, and the icons for other functions will not be
displayed.
*1 : In the case of the CR Console Lite, the Magnifying display image is optional function.
For details, refer to [
4.4 Using Utility (QA Image Controller Buttons)
].
(4) Outputting Images
Outputs images for the study currently displayed now. Select
. The output conditions follows
the setting of the “Delivery” folder. For details of the operations, refer to [
4.5 Outputting Images
].
(5) Completing Image Edit
Completes image information edit. Select
.
For details of the operations, refer to [
4.6 Completing Image Edit
].
(6) Image Edit Function Authentication button
Authenticates the operational authority of functions that change the FCR image processing
parameters. Alternately selects the display or non-display of the FCR image processing parameter-
changing buttons. When the setting box, such as the S value/L value setting box, is displayed from a
palette, operational authority cannot be changed.
Whether an operation image edit function authentication button is displayed or not depends on the
setting made at the time of installation.
Perform user authentication
:
When you select
, the “Enter Password” window appears.
After entering the password, select
.
: Cancels operation.