6-50
SECTION 6
OPTIONS
8717093
AMPS Cellular
Setup Parameters
Pass/Fall Indicator
Print-Out
Handoff Start
Handoff end
Handoff Increment
Power Lev Tolerance
RF Freq Err Tolerance
SAT Freq Tolerance
SAT Dev Tolerance
ST Dev Tolerance
Flash Dur Tolerance
13 %
10 %
1 Hz
1.00 kHz
15 %
50
300
1
DISABLED
ENABLED
1 HzST Freq Tolerance
50 %
1
2
3
4
5
6
7
8
9
10
1112
1. Pass/Fail Indicator
Displays status of Pass/Fail Indicator on Automatic Test Printout. Displays
ENABLED or DISABLED.
2. Print-Out
Enables or disables print-out of Automatic Test Output. Requires printer
connection to RS-232 Connector. Refer to Section 5 for configuring RS-232
Connector.
3. Handoff Start
First channel of Handoff Test sequence. Handoff Test starts with Handoff Start
channel, increases by Handoff Increment (5), performing Handoff Test at each
channel and ends with Handoff End (4) channel. Range is 1 to 1023. Select
using DATA ENTRY Keys. Press ENTER Key to complete edit.
4. Handoff End
Last channel of Handoff Test sequence. Handoff Test starts with Handoff Start
(3) channel, increments by the Handoff Increment (5) performing Handoff Test at
each channel and ends with Handoff End channel. Range is 1 to 1023. Select
using DATA ENTRY Keys. Press ENTER Key to complete edit.
5. Handoff Increment
Increment Number for Handoff Test. Handoff Test starts with Handoff Start (3)
channel, increments by Handoff Increment, performing Handoff Test at each
channel and ends with the Handoff End (4) channel. Range is 1 to 1023. Select
using DATA ENTRY Keys. Press ENTER Key to complete edit.