6-87
SECTION 6
OPTIONS
24. Subaud dev:
This field is active when the selected source, DATA, is modulated. This controls
the deviation of the sub-audible data used during established call processing with
the EDACS Radio. Value range: 0.00 to 100.0 kHz.
Should be
≤
Deviation (20).
The recommended default value is 0.70 kHz (0.35 kHz for narrow band).
25. Code:
This field is active when the selected source, DTMF, is modulated. This allows
the entry of desired DTMF digits to use for DTMF source generation.
26. Sinad:
This field is used to select a meter to read signals from the Audio/Data/SINAD
input baseband jack. The measurements appear to the right of this field. Valid
selections:
O
T
e
n
When baseband measurements are made, all receiver
measurements including the DSP data decoding are disabled due
to hardware limitations. Therefore, an EDACS Radio-initiated
call cannot be made with any baseband measurements enabled.
The Radio-initiated call must be made first, then the baseband
measurements may be enabled during the call. Calling the Radio
may be done at any time.