Section 3: Functions and features Model 2651A High Power System SourceMeter® Instrument Reference Manual
3-36 2651A-901-01 Rev. A / March 2011
Figure 44: Remote trigger model: Asynchronous mode
End
sweep
action
Trigger layer
Idle
smuX.trigger.initiate()
Configure and enable sweep source or sweep
measure actions, then send:
Arm event overrun
Arm layer
No
Yes
Arm
event
detector
Another
arm?
Measure event overrun
Measure
action
Measure
event
detector*
Measure
Source event overrun
End pulse event overrun
No
Yes
smuX.trigger.PULSE_COMPLETE_EVENT_ID
Pulse complete event
Source
action
Another
trigger?
End
pulse
event
detector
Source
event
detector
Source
End
pulse
action
*
* If the source loop has
completed its End sweep
action, the measure loop
will terminate unless the
Measure action block is
actively measuring. If this
is the case, the active
measurement is allowed
to complete before
returning to the arm layer.
smuX.trigger.source.stimulus
smuX.trigger.endpulse.stimulus
smuX.trigger.measure.stimulus
smuX.trigger.arm.stimulus
smuX.trigger.SWEEPING_EVENT_ID
Sweeping event
smuX.trigger.IDLE_EVENT_ID
Idle event
smuX.trigger.SWEEP_COMPLETE_EVENT_ID
Sweep complete event
smuX.trigger.ARMED_EVENT_ID
Armed event
smuX.trigger.SOURCE_COMPLETE_EVENT_ID
Source complete event
smuX.trigger.MEASURE_COMPLETE_EVENT_ID
Measure complete event

Table of Contents

Other manuals for Keithley 2651A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2651A and is the answer not in the manual?

Keithley 2651A Specifications

General IconGeneral
Maximum Voltage40 V
Current Range0 to 3 A
Resolution6.5 digits
Resolution (Current)100 pA
Minimum Current Resolution100 pA
Power300 W
Display5.7-inch color LCD
TypeSourceMeter
Communication InterfacesUSB, GPIB, Ethernet
InterfacesGPIB, USB, Ethernet
Accuracy (Voltage)0.02% + 3 mV
Resolution (Voltage)100 nV
Minimum Voltage Resolution100 nV

Related product manuals