250 Keysight InfiniiVision HD3-Series Oscilloscopes User's Guide
20 Waveform Generator
Waveform
Type
Characteristics Frequency
Range
Max.
Amplitude
(High-Z)
1
Offset Range (High-Z)
1
Arbitrary Use the Frequency/Period,
Amplitude/High-Level, and
Offset/Low-Level fields to set the arbitrary
waveform signal parameters.
Select Edit Waveform... to define the
arbitrary waveform shape. See "To edit
arbitrary waveforms" on page 254.
300 mHz to
5 MHz
2 mVpp to
10 Vpp
When amplitude <50 mV:
±(800 mV - 0.5 * amplitude)
When amplitude 50 mV:
±(8 V - 0.5 * amplitude)
Sine Use the Frequency/Period,
Amplitude/High-Level, and
Offset/Low-Level fields to set the sine signal
parameters.
10 mHz to
100 MHz
2 mVpp to
10 Vpp
When amplitude <50 mV:
±(800 mV - 0.5 * amplitude)
When amplitude 50 mV:
±(8 V - 0.5 * amplitude)
Square Use the Frequency/Period,
Amplitude/High-Level, Offset/Low-Level,
and Duty Cycle fields to set the square wave
signal parameters.
The duty cycle can be adjusted from 20% to
80%.
10 mHz to
50 MHz
2 mVpp to
10 Vpp
When amplitude <50 mV:
±(800 mV - 0.5 * amplitude)
When amplitude 50 mV:
±(8 V - 0.5 * amplitude)