Waveform Generator 20
Keysight InfiniiVision HD3-Series Oscilloscopes User's Guide 251
Ramp Use the Frequency/Period,
Amplitude/High-Level, Offset/Low-Level,
and Symmetry fields to set the ramp signal
parameters.
Symmetry represents the amount of time per
cycle that the ramp waveform is rising and
can be adjusted from 0% to 100%.
300 mHz to
5 MHz
2 mVpp to
10 Vpp
When amplitude <50 mV:
±(800 mV - 0.5 * amplitude)
When amplitude 50 mV:
±(8 V - 0.5 * amplitude)
Pulse Use the Frequency/Period,
Amplitude/High-Level, Offset/Low-Level,
and Width/Width Fine fields to set the pulse
signal parameters.
For frequencies of 101.6 Hz and lower, the
minimum pulse width is 300 ns. For
frequencies of 6 kHz and higher, the
minimum pulse width is 5 ns. For
frequencies between 101.6 Hz and 6 kHz,
the minimum pulse width goes linearly from
300 ns to 5 ns. The pulse width can be
adjusted from the minimum to the period
minus the minimum.
10 mHz to
50 MHz.
2 mVpp to
10 Vpp
When amplitude <50 mV:
±(800 mV - 0.5 * amplitude)
When amplitude 50 mV:
±(8 V - 0.5 * amplitude)
DC Use the Offset field to set the DC level.
Use the DC Mode drop-down list to select
Precise levels (with a ±1.00 V range) or Wide
range (of ±8.00 V).
n/a n/a ±8.00 V or ±1.00 V,
depending on the DC Mode
selection
Noise Use the Amplitude/High-Level and
Offset/Low-Level to set the noise signal
parameters.
n/a 2 mVpp to
10 Vpp
When amplitude <50 mV:
±(800 mV - 0.5 * amplitude)
When amplitude 50 mV:
±(8 V - 0.5 * amplitude)
Sine
Cardinal
Use the Frequency/Period, Amplitude, and
Offset fields to set the sinc signal
parameters.
300 mHz to
5 MHz
2 mVpp to
10 Vpp
When amplitude <50 mV:
±(400 mV - 0.25 *
amplitude)
When amplitude 50 mV:
±(4 V - 0.25 * amplitude)
Exponential
Rise
Use the Frequency/Period,
Amplitude/High-Level, and
Offset/Low-Level fields to set the
exponential rise signal parameters.
300 mHz to
5 MHz
2 mVpp to
10 Vpp
When amplitude <50 mV:
±(800 mV - 0.5 * amplitude)
When amplitude 50 mV:
±(8 V - 0.5 * amplitude)
Exponential
Fall
Use the Frequency/Period,
Amplitude/High-Level, and
Offset/Low-Level fields to set the
exponential fall signal parameters.
300 mHz to
5 MHz
2 mVpp to
10 Vpp
When amplitude <50 mV:
±(800 mV - 0.5 * amplitude)
When amplitude 50 mV:
±(8 V - 0.5 * amplitude)
Waveform
Type
Characteristics Frequency
Range
Max.
Amplitude
(High-Z)
1
Offset Range (High-Z)
1