252 Keysight InfiniiVision HD3-Series Oscilloscopes User's Guide
20 Waveform Generator
3 Depending on the selected waveform type, use the remaining controls in the
Waveform Generator dialog box to set the waveform's characteristics.
X Input — Select between entering the Frequency or Period of the output
waveform. Use the Frequency or Period field to enter the value.
Y Input — Select between entering Amplitude/Offset or High/Low values to
specify the output waveform's voltage swing. Use the Amplitude and Offset or
Low-Level and high-Level fields to enter the values.
Logic Presets — Sets the output voltage to TTL, CMOS (5.0V), CMOS (3.3V),
CMOS (2.5V), or ECL compatible Low and High levels.
Add Noise — Specifies the amount of white noise to add to the waveform
generator output.
Cardiac Use the Frequency/Period, Amplitude, and
Offset fields to set the cardiac signal
parameters.
300 mHz to
5 MHz
2 mVpp to
10 Vpp
When amplitude <50 mV:
±(400 mV - 0.25 *
amplitude)
When amplitude 50 mV:
±(4 V - 0.25 * amplitude)
Gaussian
Pulse
Use the Frequency/Period, Amplitude, and
Offset fields to set the Gaussian pulse signal
parameters.
300 mHz to
5 MHz
2 mVpp to
8 Vpp
When amplitude <50 mV:
±(400 mV - 0.25 *
amplitude)
When amplitude 50 mV:
±(4 V - 0.25 * amplitude)
1
When the output load is 50 , these values are halved.
Waveform
Type
Characteristics Frequency
Range
Max.
Amplitude
(High-Z)
1
Offset Range (High-Z)
1
Logic Preset Low level High level
TTL 0 V +5 V (or a TTL-compatible
high level if +5 V cannot
be reached)
CMOS (5.0V) 0 V +5 V
CMOS (3.3V) 0 V +3.3 V
CMOS (2.5V) 0 V +2.5 V
ECL -1.7 V -0.9 V

Table of Contents