Gocator Line Profile Sensors: User Manual
Gocator Device Files • 362
Trigger
The Trigger element contains settings related to trigger source, speed, and encoder resolution.
Element Type Description
Source 32s
Trigger source:
0 – Time
1 – Encoder
2 – Digital Input
3 – Software
Source.options 32s (CSV) List of available source options.
Units 32s
Sensor triggering units when source is not clock or encoder:
0 – Time
1 – Encoder
FrameRate 64f Frame rate for time trigger (Hz).
FrameRate.min 64f Minimum frame rate (Hz).
FrameRate.max 64f Maximum frame rate (Hz).
FrameRate.maxSource 32s
Source of maximum frame rate limit:
0 – Imager
1 – Surface generation
MaxFrameRateEnabled Bool Enables maximum frame rate (ignores FrameRate).
EncoderSpacing 64f Encoder spacing for encoder trigger (mm).
EncoderSpacing.min 64f Minimum encoder spacing (mm).
EncoderSpacing.max 64f Maximum encoder spacing (mm).
EncoderSpacing.minSource 32s
Source of minimum encoder spacing:
0 – Resolution
1 – Surface generation
EncoderSpacing.used Bool Whether or not this parameter is configurable.
EncoderTriggerMode 32s
Encoder triggering mode:
0 – Tracking backward
1 – Bidirectional
2 – Ignore backward
Delay 64f Trigger delay (µs or mm).
Delay.min 64f Minimum trigger delay (µs or mm).
Delay.max 64f Maximum trigger delay (µs or mm).
GateEnabled Bool Enables digital input gating.
GateEnabled.used Bool True if this parameter can be configured.
GateEnabled.value Bool Actual value if the parameter cannot be configured.
Trigger Child Elements