Gocator Line Profile Sensors: User Manual
Protocols • 493
Field Type Offset Description
width (W) 32u 12 Number of points per profile intensity array.
xScale 32u 16 X scale (nm).
xOffset 32s 20 X offset (µm).
source 8u 24 Source
0 – Top
1 – Bottom
2 – Top Left
3 – Top Right
exposure 32u 25 Exposure (ns).
reserved[3] 8u 29 Reserved.
points[C][W] 8u 32 Intensity arrays.
Surface
Field Type Offset Description
size 32u 0 Count of bytes in message (including this field).
control 16u 4 Bit 15: Last message flag.
Bits 0-14: Message type identifier. For this message, set to 8.
attributeSize 16u 6 Size of attributes, in bytes (min: 40, current: 48).
length (L) 32u 8 Surface length (rows).
length (W) 32u 12 Surface width (columns).
xScale 32u 16 X scale (nm).
yScale 32u 20 Y scale (nm).
zScale 32u 24 Z scale (nm).
xOffset 32s 28 X offset (µm).
yOffset 32s 32 Y offset (µm).
zOffset 32s 36 Z offset (µm).
source 8u 40 Source
0 – Top
1 – Bottom
2 – Top Left
3 – Top Right
exposure 32u 41 Exposure (ns).
reserved[7] 8u 45 Reserved.
streamStep 32s 52
Data stream step number. For a surface, values are:
3 – surface stream step
8 – tool data stream step

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LMI Technologies Master 800 and is the answer not in the manual?

LMI Technologies Master 800 Specifications

General IconGeneral
BrandLMI Technologies
ModelMaster 800
CategoryAccessories
LanguageEnglish

Related product manuals