EasyManua.ls Logo

LORAD AFFINITY Series - X-ray and Light Field Check and Alignment Procedures

LORAD AFFINITY Series
214 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-44 P/N 9-500-0255
Service Manual
Chapter 4: Calibrations and Performance Checks
X-ray and Light Field Check and Alignment Procedures
7.0 X-ray and Light Field Check and Alignment Procedures
The following checks and adjustments are used to ensure that the alignment of the x-ray field
and light field are within the specifications set forth by the FDA, 21CFR, and the
recommendations by the ACR/CDC.
X-ray field alignment is the adjustment of the built-in and fixed apertures so that the x-ray
field is collimated within specified limits. The light field is adjusted to be congruent with the
x-ray field within 2% of SID.
X-ray field adjustments are performed in two stages: coarse adjustment (using a light-
emitting x-ray screen) and fine adjustment (using x-ray films). The coarse adjustment
procedures employ an X-ray Beam Alignment Template that is to be used with a separate 14
x 17 cm x-ray cassette screen.
7.1 X-ray Beam Alignment Template
The translucent X-ray Beam Alignment Template (LORAD P/N 3-405-8010), shown in Figure
4-10 must be positioned on the units image receptor support on top of a 14 x 17 cm cassette
screen (Lanex Med). The visible borders of the x-ray field can be compared with graduations
on the template to determine coarse x-ray field alignment
The template contains minimum and maximum borders for the left, right, and rear field
edges of the 18 x 24 cm format and the 24 x 30 cm format. For the chest wall edge, a
maximum limit (6.5 mm) is provided to adjust for chest wall edge overlap.
Always position the X-ray Beam
Alignment Template accurately
on the image receptor support
by performing the following:.
1. Place the screen (facing up)
onto the support device and
center it.
2. Move the screen forward
(away from the C-arm)
approximately 1.5". Lay the
template on top of the
screen.
3. Butt the rear edge of the
template against the C-arm
to place the image receptor
tray edge line directly
above the edge of the
image receptor support
device.
4. Center the template left-to-
right until the cut in edges
(at the rear of the template)
are set flush with the sides of the image receptor support.
Figure 4-10: X-ray Beam Alignment Template

Table of Contents