P/N 9-500-0255 6-39
Service Manual
Chapter 6: Maintenance—Remove and Replace Procedures
C-Arm Components—Remove and Replace
D. Press S2, verify the lamp and mirror assemblies move out of field.
E. Press S4, verify the lamp and mirror assemblies move back in.
F. Press S3, verify the filter select mechanism changes position.
G. Press S4, verify the entire collimator assembly cycles in and out of field.
Table 6-1: Tubehead Control Board Push Button Switches.
8. Turn power OFF, replace all covers and return to service.
Switch Function
S1 Auto Aperture Select
S2 Light Field Lamp Assy/Motorized Mirror Assy Select
S3 Filter Select
S4 Collimator Assy Select