5.3 Explanation of Each Program Message
BP4610
5-8
5.3 Explanation of Each Program Message
The details of the program messages are described below. Commands are listed in alphabetical order.
5.3.1 Basic settings
ACCU/ACCU?
Description: Sets the superimposed AC current.
Parameter: Current [Ap-p]
0.00 Ap-p to 30.00 Ap-p
Setting example: Setting the superimposed AC current to 20.00 Ap-p
ACCU 20
Caution: Settable in units of 0.01 Ap-p.
ACVL/ACVL?
Description: Sets the superimposed AC voltage.
Parameter: Voltage [Vp-p]
0.0 Vp-p to 120.0 Vp-p
Setting example: Setting the superimposed AC voltage to 100 Vp-p
ACVL 100
Caution: Settable in units of 0.1 Vp-p.
CURR/CURR?
Description: Sets the DC current.
Parameter: Current [A]
−10.000 A to 10.000 A
Setting example: Setting the DC current to 5.000 A
CURR 5
Caution: Settable in units of 0.001 A.
FREQ/FREQ?
Description: Sets the superimposed AC frequency.
Parameter: Frequency [Hz]
1.0 Hz to 100000.0 Hz
Setting example: Setting the superimposed AC frequency to 70000 Hz
FREQ 70000
Caution: Settable in units of 0.1 Hz.