5.3 Explanation of Each Program Message
BP4610
5-15
5.3.4 Measurement related
MACC?
Description: Returns the AC current measurement value.
Response format: 0.0 Ap-p to 40.0 Ap-p (returns 99.9 Ap-p in the case of measurement over-
range)
Caution: Measurement is performed in units of 0.1 Ap-p.
MACV?
Description: Returns the AC voltage measurement value.
Response format: 0 Vp-p to 400 Vp-p (999 Vp-p is returned in the case of measurement over-
range)
Caution: Measurement is performed in units of 1 Vp-p.
MCUR?
Description: Returns the DC current measurement value.
Response format: −20.00 A to 20.00 A (99.99 A is returned in the case of measurement over-
range)
Caution: Measurement is performed in units of 0.01 A.
MVLT?
Description: Returns to the DC voltage measurement value.
Response format: −200.0 V to 200.0 V (999.9 V is returned in the case of measurement over-
range)
Caution: Measurement is performed in units of 0.1 V.
5.3.5 Arbitrary waveform memory
The settings performed with ARBM and ARBT are done for the arbitrary waveform memory specified
with ARBN. If no specification is made with ARBN, arbitrary waveform memory 2 is used as the
default. To control the arbitrary waveform saved to another memory, first specify the memory number
with ARBN.
ARBM
Description: Controls arbitrary waveform memory.
Parameter: 0: Clear
Setting example: Clears arbitrary waveform.
ARBM 0