8.4 Signal Sources
BP4610
8-5
8.4 Signal Sources
The signal source can be selected from among internal signal source, external signal source, and
internal signal source + external signal source.
Internal signal source
• Current
Amplitude setting range: ±115 V (0.01 V resolution) (CV mode)
±10 A (0.001 A resolution) (CC mode)
Amplitude setting accuracy: ±(|0.5% of setting value| + 0.2 V) (CV mode,
AC 0 V, 23 Ω resistance load, DC only, 23 ±5°C)
±(|0.5% of setting value| + 0.03 A) (CC mode,
AC 0 A, DC only, 23 ±5°C)
• Superimposed AC
Waveform: Sine, square, arbitrary (16 types)
Frequency setting range: 1 Hz to 100 kHz (0.1 Hz resolution)
Frequency accuracy: ±100 ppm
Amplitude setting range: 0 to 120 Vp-p (0.1 Vp-p resolution) (CV mode)
0 to 30 Ap-p (0.01 Ap-p resolution) (CC mode)
Amplitude setting accuracy: ±(0.5% of setting value + 0.2 Vp-p) (CV mode,
DC 0 V, AC only 500 Hz square wave, 23 ±5°C)
±(0.5% of setting value + 0.03 Ap-p) (CC mode,
DC 0 A, AC only 500 Hz sine wave, 4 Ω resistance load,
23 ±5°C)
External signal input
Gain: 100× (100 V/1 V), in phase (CV mode)
10× (10 A/1 V), in phase (CC mode)
Gain accuracy: ±5% (1 kHz)
Input impedance: 10 kΩ
Non-destructive max. input voltage: ±5 V
Input terminal: BNC connector (front panel)
Frequency range: DC to 200 kHz