Others - Sequencer
8 - 9
Table 1:
Display of calculated
value in measurement
table
Calculated value
Both the items to select from in this combo box and the corresponding
display at the measurement table below depend on the selected
measured quantities.
In this combo box you determine whether the measurement table
displays trigger signal characteristic / elapsed state time until trigger
signal occurrence, ratio and phase angle difference, impedance Z or
resistance R, real power P, apparent power S or the power factor cos
ϕ.
Measured
quantities
Display of calculated value in measurement table
(m1) (m2)
Bin/Time If selected, the column "BinIn" displays the characteristic of
the trigger signal at the binary input BinIn, and the column
"s" the elapsed state time until the occurrence of the trigger
signal in seconds.
4 different characteristics are possible at BinIn:
Closed: potential-free contact between BinIn+ and BinIn–
closed.
Open: potential-free contact between BinIn+ and BinIn–
open.
Come:
trigger signal with rising edge occurred at BinIn.
Go:
trigger signal with falling edge occurred at BinIn
V V Ratio:1, Ratio:5, Diff:
Ratio m1/m2 and phase angle
ϕm1 - ϕm2 (if phase angles
are available; otherwise "n/a"), and differences
ΔU and
Δϕ
U
.
I I Ratio:1, Ratio:5, Diff:
Ratio m1/m2 and phase angle
ϕm1 - ϕm2 (if phase angles
are available; otherwise "n/a"), and differences
ΔI and Δϕ
I
.