CPC 100 V1.41
2 - 40
The logging level specifies the occurrences to be recorded in the log file:
The default setting is "Warning".
If the log file gets quite long, use either the scroll bar or the buttons P
AGE UP /
P
AGE DOWN to navigate up and down in the list of logged occurrences.
C
LEAR LOG deletes all log file entries and starts recording occurrences anew.
Logging level: records
None nothing
Error internal errors, hardware errors, operational errors,
overload conditions etc.
Warning uncritical errors such as over-temperature, malfunction
etc.
Info additional operational information about CPC 100.
This setting is suggested if the operator suspects, for
example, an incorrect functioning or an operational
malfunction of CPC 100. This setting records every
relevant operational activity of CPC 100.
The log file can then be sent to OMICRON for
diagnosis.
Note: The logging levels are hierarchical, that is, if logging level "Warning" is
set, "Errors" are also recorded, or if "Info" is set, the logging levels
"Warning" and "Error" also apply.