Getting Started with Quick
3 - 9
Power output value
Current (I) or voltage (V) value, depending on the
specified range (see above).
Any value between minimum and maximum as well as
the explicit value 0.0 is possible, even while in the "on"
state.
The respective minimum and maximum values depend
on the range and, for voltage ranges, also on the
frequency. For frequencies smaller than 50Hz, and down
to the minimum frequency of 15Hz, the possible
maximum voltage value decreases linear proportional to
the frequency value (refer to Figure 3, ”Typical maximum
output voltage of 2kV AC output depending on frequency
and power” on page 16-14).
Current ranges do not have such a dependency.
Frequency / Phase angle
Frequency value in Hz or - if "Sync w/ V1AC" is selected
(see below) - phase angle in degrees. For DC ranges this
data entry field is disabled.
The frequency can be set within a range of
15.00 ... 400.00Hz, even while in the "on" state.
The phase angle is displayed in a range of
-180.0 ... +180.0°, and can also be changed at any time.
If this range is exceeded, the phase angle will adapt itself
automatically, that is, change from -180.0 to +180.0°.

Table of Contents

Other manuals for Omicron CPC 100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omicron CPC 100 and is the answer not in the manual?

Omicron CPC 100 Specifications

General IconGeneral
BrandOmicron
ModelCPC 100
CategoryTest Equipment
LanguageEnglish

Related product manuals