LIFEPAK 20/20e Defibrillator/Monitor Troubleshooting
Service Code Table (continued)
(Continued on next page)
A01 System PCB A02 PP PCB Coin Battery
Code Service Code Description Corrective Action
501d RTC out of sync 1. Replace the coin cell battery
2. Replace the A01 System PCB
501e System software execution error Reload the device software*
501f System software read error Reload the device software*
5020 System software write error Reload the device software*
5021 System software error Reload the device software*
5022 System software exception code Reload the device software*
5023 System software exception code Reload the device software*
5024 System software exception code Reload the device software*
5026 NVRAM low battery interrupt Replace the coin cell battery
5027 System USB did not initialize 1. Clear error and perform PIP
2. Replace the A01 System PCB
5028 PP USB did not initialize 1. Clear error and perform PIP
2. Replace the A02 PP PCB
6-23