LIFEPAK 20/20e Defibrillator/Monitor Troubleshooting
Service Code Table (continued)
A02 PP PCB A10 SpO2 Module
Code Service Code Description Corrective Action
6002 PP program corrupted Reload the device software*
6003 PP program not found Reload the device software*
6004 PP boot-up error Reload the device software*
6009 No PP data Reload the device software*
600c SpO2 misconfigured 1. Check configuration
2. Reload the device software*
600e PP reset 1. Reload the device software*
2. Replace the A02 PP PCB
600f OEM configuration error 1. Check configuration
2. Reload the device software*
6010 PP initialization error Reload the device software*
6018 Incorrect updated SpO2 image Replace the A06 OEM PCB
6019 Incorrect updated SpO2 image CRC Replace the A06 OEM PCB