LIFEPAK 20/20e Defibrillator/Monitor Replacement Procedures
Front Case (continued)
A11 Active Display
Installation (continued)
8. A11 Active Display - 3309555, route display backlight wires together around
black strain relief features. Extend excess wires to the right, straight across
the top of the display.
9. Position the A11 Active Display inside the display bracket.
10. A11 Active Display - 3205278, route the backlight wires from display along
the corner and under the hook and then back to the A08 Backlight Inverter.
11. Connect the A11 Active Display wires to the A08 Backlight Inverter at J2.
12. Place the active display bracket assembly in position in the front case. (If
replacing the Active Display, remove the clear protective cover prior to
installing it into the front case.)
Note: A08 Backlight Inverter - 3202033 must make contact with the thermal
conductive pad (201) on the front case.
13. Place the two grounding harnesses (219) onto the top two screws (173);
torque to 6.8 in-lb. Refer to Grounding Harnesses Orientation for
grounding harness placement.
(Continued on next page)
Parts List Interconnect Front Case View Rear View Main Assemblies
9-47