B(k) the Lden result (ccc – the value in dB);
Y the Ltm3 result (ccc – the value in dB);
Z the Ltm5 result (ccc – the value in dB);
L(nn) the value L of the nn statistics (ccc – the value in dB);
r the underrange flag (ccc equals to 0 or 1);
In case of results timestep other than measurement time, only VPMNR subset of results is available
Note: The value displayed on the screen during the result’s presentation will be sent out from the
instrument in the case when after the X code the nn is not given.
Note: For profiles 2 and 3 the L(nn) result is not calculated.
Note: The presented above order of the measurement results sent out by the instrument does
not depend on the order of the characters sent to the unit.
Note: In the case of Lden, the value k placed in the parenthesis after the code B, denotes the
kind of the currently measured result. The kind of the Lden result depends on the time during which the
measurements were performed (d denotes day, e denotes evening and n denotes night).
The corresponding values of k parameter and the kind of the measured Lden result are presented below:
k = 1 Ld result,
k = 2 Le result,
k = 3 Lde result,
k = 4 Ln result,
k = 5 Lnd result,
k = 6 Len result,
k = 7 Lden result.
Example: After sending to the instrument the following string:
#2,1,T?,V?,B?,P?,M?,R?,L50?;
one should receive the answer given below:
#2,1,T3,V0,P66.91,M64.55,R61.70,B(2)66.70,L(50)54.95;
The codes of the results in the case of SOUND DOSIMETER mode are defined as follows:
T time of the measurement (ccc – value in seconds);
V the overload flag (ccc equals to 0 or 1);
P the PEAK value (ccc – the value in dB);
M the MAX value (ccc – the value in dB);
N the MIN value (ccc – the value in dB);